How to Use the Cotton Tip Probes With the QT+Updated a month ago


Using the Cotton-tip probes and Precision Wand Tips with the MyoLift QT+. 

* Available only for Premium Membership Users * 

  1. Connect your MyoLift QT+ lead wire to your MyoLift QT+ device.
  2. Open the Myolift QT app and select the “Custom Treatment” option, located in the “Sessions” menu of your MyolIft QT app.
  3. In the Customer Treatment, select the time duration for the STRETCH waveform and the LIFT waveform.
  4. Next, select the applicator: You can select either the Mask or the Gloves option for this treatment.  Please note that the Probe option is not compatible with this treatment.
  5. Lastly, select the intensity you would like to use and begin your treatment. 



TROUBLE SHOOTING


Check the Lead Wire:  

  1. Inspect the lead wire for any signs of damage.
  2. Make sure that the lead wire is fully connected to the device.
  3. Test the lead wire to make sure it conducts.
  4.  Start a treatment and touch the two pins at the end of the lead wire together and hold for 10 to 15 seconds.  
  5. If the conductivity bar shows the device is conducting, we can rule out any issues with the lead wire. 

Test the wands:

  1. Connect the wands to the lead wire
  2. Start a treatment and touch the tips of the wands together. Hold the wand together for 10 to 15 second. 
  3. If the conductivity bar shows the device is conducting, we can rule out any issues with the wands.

additional tips:

  • We recommend using the Q-tip brand only and make sure the shaft is made out of cardboard. Q-tips that have a plastic shaft will not conduct. 
  • Cut the Q-tips to 1/4 inch in length
  • Make sure you  soak the cut Q-tips in the Infused conductive solution  before you place them into the ends of the probes. 
  • Push the cotton tips down into the cotton-tip probe for a snug fit. 
  • Dip the tips into the Infused conductive solution to wet the Q-tips again, before starting your treatment. 
  • You can test touching the two q-tops together to see if there is conductivity. 


Other factors that may affect conductivity are:

Dehydration : We recommend that you drink at least 3 cups of water, 30 minutes before you begin your treatment. You can also use the 7e RePlenish Conductive Spritzer to help improve dehydration. 

Scar Tissue or Surgical Scars: Sometimes areas on the skin that have been compromised, may cause the loss of conductivity.